MRI SAFETY INFORMATION - RemeOs™

Non-clinical testing demonstrated that the RemeOs™ Screw is MR Conditional. A patient with this device can be scanned safely in an MR system under the following conditions. Failure to follow these conditions may result in injury:

Static Magnetic Field Strength (B0) 1.5T or 3.0T
RF Excitation Circularly Polarized (CP)
RF Transmit Coil Type Volume RF body coil
Operating Mode Normal Operating Mode
Maximum Whole-Body SAR 2 W/kg (Normal Operating Mode)
Maximum Head SAR 3.2W/kg
Scan Duration Under the scan conditions defined, the implant can be scanned continuously for 25 minutes. With 5 minutes break, this can be repeated twice in 60 minutes.