Non-clinical testing demonstrated that the RemeOs™ Screw is MR Conditional. A patient with this device can be scanned safely in an MR system under the following conditions. Failure to follow these conditions may result in injury:
| Static Magnetic Field Strength (B0) | 1.5T or 3.0T |
| RF Excitation | Circularly Polarized (CP) |
| RF Transmit Coil Type | Volume RF body coil |
| Operating Mode | Normal Operating Mode |
| Maximum Whole-Body SAR | 2 W/kg (Normal Operating Mode) |
| Maximum Head SAR | 3.2W/kg |
| Scan Duration | Under the scan conditions defined, the implant can be scanned continuously for 25 minutes. With 5 minutes break, this can be repeated twice in 60 minutes. |